Introduction
X-ray tomography is considered more than essential nowadays in various fields1, and combined with advanced X-ray sources, it provides more sophisticated scientific insights2. As Wang et al.3 once argued, it is desirable to achieve an automated data processing with minimum human interaction due to the demand of high-throughput experiments. Given the fact that we only get the 2D projections from the x-ray penetration to form the internal 3D structure of a sample, a number of scientists have shared their insights to get rid of impairing effects in getting a better quality of reconstruction, translation and tilt errors, for example.
They have focused on the determination of the center of rotation (CoR), and there are mainly three families of methods. The first category uses pairs of projection images taken from the reverse viewing angles (at 0–180 degrees)4. The second method evaluates the projection image from the image reconstruction by using a parameter to measure the quality of image reconstruction and to calibrate the relative offset of the rotation axis5. The last one considers the center-of-mass (CM)6. In the previous researches7,8, researchers tried to find the rotation axis (RA) by projection image rearrangement, but there was no proper method to find the RA even in 2-dimensional space9. Hence it is always useful to figure out a new method and algorithm to correct the flawed information to obtain a reconstruction of better quality.
In this article we propose an alternative approach to remove the errors, specifically translation and tilt errors by using the sinogram and the fixed point. Wherein, fixed point is defined as the point in space that can be discriminated or calculated by analysis of a projection image set. The actual RA is located on the center line of the sinusoidal trajectory of the Fixed point (FP). Furthermore, we did not try to determine the CoR; instead, we aligned the vertical center line of projection to the CoR to make it like a virtual RA, which was enabled by our algorithm.
The relative position of an object in the real space and the sinograms in 2D
A sinogram involves the information about a specific layer of an object and accumulates the projection shadows taken from each projection angle to build a reconstruction of the layer. This sinogram will later be transformed into a reconstruction through the inverse Radon transform. Apparently, an errorless reconstruction will be obtained only with an ideal sinogram or an ideal projection set, and a necessary condition for the ideal sinogram is that its pattern is to be changed sequentially depending on the projection angles.
Assertion 1 When the location of an object is changed on the stage, the sinogram pattern will be also changed while the projection shadow from an angle remains the same.
There might be slight differences in the projection shadow due to the limit of image digitalizing techniques; this can be approximated by a linear interpolation. Where a projection shadow lies on the sinogram is closely related to where an object lies in the real space; this relationship should be first investigated before we further use sinograms for our analysis. The object on the rotating stage is known as to rotate around the rotation axis (RA) and it means that any single point of the object has its own circular trajectory, which will be expressed as a sinusoidal function in the sinogram. The point on the RA will be the straight line across the center of the sinogram in ideal condition. Our hypothesis is that if we have the projection set of a specific object and if we know the ideal sinogram pattern of this object, we can modify the actual sinogram pattern to meet the ideal one and consequently get the errorless reconstruction. By modifying the sinogram, we can make the same effect just like moving the specimen in real space from where it was to the center on the stage, while the recorded image is preserved. This saves the current efforts of determining the CoR and setting the specimen on the RA, and the simple modification of sinogram will yield the same result. (see Supplementary Section 1).
2D Correction Algorithm: Virtual Focusing Method using FP
In the actual settings of computed tomography (CT) system, the translation error is not only either a horizontal or vertical error but also a combination of both. Thus, an optimal solution for the orthogonal translation error correction (see Supplementary Section 2.2) is not enough in general. The horizontal shift of an object is hard to be detected on the projection image and so, difficult to be calculated. This leads us to a need of approaching this issue by adopting a fairly new concept and a point of view. Knowing that the errors in the actual settings are complicated and often a mixture of different types of errors, one should start with understanding and defining the relationship between the real space and the sinogram, which is actually a projected and reconstructed version of real space, so as to use it to draw a general and automated method of correction in advance.
Ideally, the center line of a sinogram is made up of the projection of RA. Any single point in the real space will move on a circular trajectory rotating around the RA and this will correspond to a function on the sinogram. Let’s say there is a point p and it moves on a circular trajectory, then this point p will be transformed into a curve on the sinogram and represented by the sinusoidal function Tr,φ.
Assertion 2 When an object rotates around RA, the circular trajectory for a point p of the object can be represented by a curve in the sinogram.
We defined r as the distance between the rotation axis and the point p. θ is the projecting angle, and φ is the angle between the line and the orthogonal line to the projection angle at θ = 0 (Fig. 1b). In ideal cases, the center O is converted to T0,φ in the sinogram, but not in the actual sinogram.
Using Assertion 1 and 2 which were mentioned earlier in our study, we can alter an inaccurate sinogram to a correct one. Tr,φ is a function that shows how a specific point p in the real space moves on the sinogram. It means that we can keep tracing a point of the real space on the sinogram and use it further as we intend to. In other words, if a point p in the solid specimen rotates on the stage and the projected curve drawn by the movement of p for each angle is the same as the sinusoidal curve by Tr,φ in the sinogram, then the projected trajectories of other points in the specimen should satisfy the projected curves by and all points in space satisfy the Helgason–Ludwig consistency condition. Being the center of a circle, the point p on the sinogram in Fig. 1a will be always on the center of the projection shadows. It follows the curve marked as a black line on the sinogram since the specimen itself is off the center. The point p is a center-of-mass for a circular object if the object has an identical medium and acts as a fixed point to represent the same spot even when the projection angle changes. With the point p translated on the function T0,φ, that is to say, with each column in the sinogram moved so that the point p is on the center and the black lined curve is laid linearly on the center line of the sinogram (Fig. 1c), the center of an object will match exactly to the center of projection and the sinogram will become linear. When applied to a real space, this virtual translation has the same effect as we physically set the center of an object on the RA.
Nevertheless, actual objects are not always cylindrical. They rather come in much more complicated structures, and as a result, defining a fixed point on the projections is not an easy issue. We came up with an advanced physical concept called the center of attenuation (CA) to make this issue simpler. The CA is a similar idea with the center-of-mass, but is different in a way that a unit particle of an object is expressed not by the X-ray density but by the x-ray mass attenuation coefficient10 (MAC) which is measured by the unit voxel. In general, the center of mass estimated by X-ray densities in projection images can’t play a role in our FP because of the followings: the X-ray absorption for the thickness of specimen is not linear and the specimen size is larger than the charged coupled device (CCD). Our assumption for using MAC is that there is co-relationship between a unit voxel of actual object and a unit voxel of reconstruction (see Supplementary Fig. S7). For 3D, the CA calculation requires the object part in a subset of common layers. (We will discuss the common layer in Method).
Assertion 3 When the relationship between the absorption of photons and the thickness of specimen is not linear after the x-ray beam penetrates the specimen, it should be first transformed into something with linearity and then utilized to figure out the CA.
When the value of MAC is changed into the one expressed in length to calculate the CA (See Method). For the calculation of the CA, all pixel values obtained by X-ray density in projection should be expressed to the sum of MAC. For the calculation of the CA, all pixel values obtained by X-ray density in projection should be expressed to the sum of MAC. When an x-ray beam is shot to an object, the object absorbs certain amount of the energy and the rest of attenuated energy arrives on the detector. Then we get the projection image. We assumed that there exists an invariable center of MAC for a certain object, just like every object has a center of mass obtained by classical mechanics, and it is fixed on a certain spot, either inside or outside of the specimen, acting as a fixed point that does not change depending on the projection angle. This fixed point can be calculated from the projection images obtained from each angle of x-ray penetration. It will lie on the sinogram satisfying the Tr,φ function that was mentioned earlier and this is how we get an ideal sinogram pattern. Figure 2 shows how we applied this idea of CA to the image reconstruction.
Analysis of 3D Image Reconstruction from a Projection Image Set
There are only translation errors in the two dimensional projections. However, other types of error have to be taken into consideration when it comes to three dimensions. Mainly, three types of errors can be discussed; a translation error caused by the shift of a specimen, a tilt error caused when the RA is tilted, and the rotation error that is caused when a specimen spins on its own axis.
A translation error is the error that occurs when the object is moved by any chance during the beam time, and this movement can be in three directions in the 3D space. A compensation for this error can be done through just the way we did with the translation error in 2D space, bringing projected fixed points on the Tr,φ,h function (h is a height of a specific layer of projections. See Method).
In 3-dimensional space, it is important to consider tilt errors because a tilted image carries information from different layers and consequently induces a flawed image reconstruction. Thus, if we compensate the tilt error, it means that we make one layer to carry all the information about a single part of an object. To categorize and analyze tilt errors, we need to think about the object itself and the stage that the specimen is placed on. We will not call the case of tilted object an error. It is because we can make a reconstruction without any correction procedures in this case (see Supplementary Section 3.1). When it is the RA that is tilted, one should make sure that the tilt error is corrected.
In terms of the tilted RA, we can classify this into two cases of tilt errors in detail; one is of a vertical direction and the other is of a horizontal direction to the x-ray beam. The former is shown in the Fig. 3(b) and the latter is shown in the Fig. 3(c). In a vertical tilt, the polar angle increases when the azimuthal angle is either 90 or 270 degrees. On the other hands, the polar angle increases when an azimuthal angle is either 0 or 180 degrees in a parallel tilt. The azimuthal angle increases counter clockwise against the beam. The ideal RA which stands upright without any tilt error is assumed to be the reference for the polar angle.

The third and fourth pictures in (a,b and c) show the trajectories of . When the
is on the RA (orange dot in the picture), it is expressed as a dot. In other cases, it is expressed as a circular trajectory (blue circle). (a) The sinogram and the
trajectory when an object is tilted. Even though the object is tilted, the
is on the parallel line with the stage in the projection. This is quite a typical occasion where no error is found. (b) The sinogram and the
trajectory when the RA is vertically tilted. The object rotates around the RA and the line that
makes are perpendicular to the RA on the projection. (c) The sinogram, its reconstruction and the
trajectory when the RA has a parallel tilt. The pattern of the sinogram is linear around the center. However, its reconstruction is flawed (shown better in the magnified figure), since the layers are all mixed up at each angle θ. In this case, the collection of
s in the projection makes an elliptical shape, not a line. The RA is also perpendicular to the major axis of
trajectory in this case.
It is the case of a vertical tilt when the RA is tilted against the axis of whole projection and the stage is rotating with the tilted RA. If we assume that the common layer is also tilted in parallel with the stage, namely tilted by the same angle in the projection, the projection information is now about a single layer, not mixed with the ones about other layers. Hence, if we only know how much the RA is tilted, and carry out a rotation compensation on the whole projection rotating around the CA, we can find the common layer that we want to concentrate on (see Method). Contrarily, the parallel tilt carries information about more than one layer and it is hard to get an ideal sinogram pattern in this case.
Discussion and Conclusion
In this study, we categorized the errors that possibly take place during the beam time. Although there certainly is a limit to what extent we can correct the reconstruction in each case, we can modify most of the errors using FP and enable the improvement of the reconstruction quality. We call the reconstruction as the ideally focused reconstruction from now, which is coming from ideal sinogram pattern. There are two cases for the tilt errors. When it is the vertical tilt, we get the ideally focused reconstruction. Meanwhile, we will not be able to get an ideally focused reconstruction with the projection set for the parallel tilt since the information of different layers are mixed up. The tilt error is mostly a mixture of both vertical and parallel tilts; the optimized projection set in which the vertical error is compensated is the best deal in this case. For the rotation error, it seems impossible to reconstruct images appropriately because we hardly get the complete projection set.
The CA we suggested in this study will function as the fixed point that is one of the intrinsic factors of an object. It works as an invariable point inside (sometimes outside) of an object which does not change even when the object is translated, or the RA is tilted. Nonetheless, we have to calculate this CA from the projected image so as to utilize it, which results in one limitation that the intensity variation of the beam reaching each cell of the CCD in the formula for should be linearly proportional to the length of the specimen. That is, the value αk of each cell which consists of the image from the projection should have certain linear relationship with the length of the specimen, or at least should be changeable to have linear relationship (see Supplementary Section 5 and Fig. 4). Particularly in soft X-ray tomography (SXT), there is a linear relationship11 and a unique linear absorption coefficient measurement is given12,13. Scientists have successfully used imaging of SXT for 3D images of various cell types with isotropic resolution12,14,15,16,17,18,19. They have tried to automate the tracking process of fiducial markers through a stack of projection images20,21,22,23,24,25. The trajectories of these fiducial markers follow Tr,φ,h function; the fiducial markers can also be used as fixed points.
(a) The white parts in the projection image represent the high density areas. The marked rectangle represents a common layer including high density areas. (b) This reconstruction was made by ideal sinogram pattern, but the density of the grain depends on the position of the reconstructed object. (c) Two magnified images in the reconstruction of (b).
To obtain a well reconstruction, we need to get a stable x-ray density against the object regardless of projecting angles. In reality though, we sometimes get the different x-ray density according to the projecting angle changes, and then, the value of A(see Method) also should be changed for the formula applying . It would be difficult to generally use
for the image reconstruction if the location of CA depends on the changes of x-ray density and the A value. The changes of x-ray density (according to the angle) will appear in forms of the shadow intensity in the sinogram. Nonetheless, it won’t affect the pattern of the sinogram. Figure 2d and e proves to us that the location of CA does not vary despite the changes of A value against each θ caused by the different x-ray density. We may be able to get a better image if we mathematically modify the x-ray density in order to even up the A value. Furthermore, if we can find the minimum value of A preserving the linearity, we can obtain sufficiently large value of A by mathematical modification. For the total amount of radiation dose, we can get the better reconstructions by increasing the total number of projection images with low dose X-ray. When the specimen is longer than the CCD, its whole image is not projected and consequently, there would be a limitation in applying the CA. We expect to discover another solution even in those cases if we can figure out the common layer on which the object is commonly projected and apply the CA on it.
In spite of the benefits of categorization, the errors often come about not as we categorized, but as a mixture of different types of errors. It is not possible to perfectly restore the image even with the CA and the Tr,φ,h function in those cases. The application will be much more extended if we can discover more fixed points besides CA and we can utilize each of them to the Tr,φ,h function. Although further researches should be proceeded, we simply showed the example of this idea in Fig. 5. If there is an area whose boundary is definite and x-ray impermeability is comparatively high, it will be distinguished in most of projections. We might be able to use its center as another fixed point. We searched for the trajectory of the fixed point in the sinogram of Fig. 5(a), and corrected a vertical translation error as Fig. 5(b) (See supplementary Section 2.2). The fixed point was transformed into the function T0,φ,h in Fig. 5(c). Here, we see that it provided us with a clearer image when the T0,φ,h function was applied to the fixed point. It is expected most of the errors will be corrected when the fixed point and the CA applied to the Tr,φ,h function in case of mixed errors.
The third pictures in (a,b, and c) are the magnified images of the yellow frames of the corresponding reconstruction s in the middle. (a) From the analysis of this sinogram, the specific fn(t) value was found at the discontinued spot; it was expected that a vertical translation error arose between the 360th and the 361th projections. For the middle panel, we used the RA of part B to minimize the error from two distinct RAs. (b) We corrected a vertical translation error (see Supplementary Section 2.2) to the part A of the sinogram (a), which brought us a sinogram with the continuity and a better reconstruction. The area with high x-ray impermeability is moving along inside the sinogram showing a sinusoidal pattern (black arrows in left panel). In its reconstruction, this area is placed in the upper right part (white arrow) in the middle panel. (c) The center of the x-ray impermeable material is set as the fixed point and it is applied to the function T0,φ representing the virtual rotational axis of the sinogram; the FP is now shifted on the line across the center in the sinogram (black arrows). It is placed on the center of the image in the reconstruction image (white arrow).
Additional researches are needed because of the limits of applying the CA to real images. We will continue to seek for the correct algorithm in applying the CA and find out more traits that can be used as the fixed points. This endeavor will contribute to getting better reconstructions even when the various errors coincide, when the whole image is not obtained because the specimen outlies the CCD, and when the parallel tilt error is included.
Method
The calculation of CA is very similar with the one for center-of-mass obtained by classical mechanics. First, a virtual rectangular coordinate and unit cubes are adopted in real space and each vertex is the whole number on the coordinate. Let’s assume that the specimen placed in the coordinate is composed of n unit cubes. The k-th cube among the n cubes of the object has a mass attenuation coefficient μ/ρ (MAC) for the given unit cube related to voxel length on its designated location and it is represented as ai, which can be linearly calculated through a series of calculation processes. Other coefficients related to X-ray transmittance which might represent the specificity of material, such as mass energy-absorption coefficients, can be used for the same purpose.
Below is the formulation we used to mathematically identify the location of CA on the coordinate.
αk is , and c satisfy that an unit voxel density for the given initial energy has the same value of αk. A is the total sum of αk and a constant value. A is defined in a subset of a whole common layer set (this will be discussed in 3D) and n is the total grid number of the object in a subset of common layers of the real space.
indicates the center of k-th grid.
is one of the fixed points in real space and particularly significant in that it will be always projected and we can trace it later on the projections. So,
, a projected position of
can be calculated in the projection image as below
Here we see that , and pij is the modified (i, j) pixel value of the projected specimen in the 2-dimensional projection and
is the center of the pixel. As the αk is calculated by MAC, the pij also should be linearly expressed and modified by the sum of the related αk.
In the projection image, the MAC of the area except for the computational domain for the object shadow should be zero in the ideal status. If it is not zero, might not be able to act as the fixed point. So we need to modify the attenuation value of this area to have at least the average of zero when it is not exactly zero, ensuring the same amount of the value is added to or subtracted from the pij area.
in real space becomes a specific point that we know and it is projected to be
and expressed as the Tr,φ function on the sinogram. The center of projection is always the center line of the sinogram. Therefore, when
from each angle of x-ray beam is translated onto the center line T0,φ of the sinogram, it can be considered the same as the fixed point of an object is placed on the RA. In other words, we are able to get rid of all the translation errors in the 2-dimensional space by translating
of each projection on the Tr,φ function. Since
of each projection is one of the projected fixed points, all fixed points can be translated on Tr,φ and expressed in the sinogram. For actual samples, the center line of the sinogram doesn’t represent a RA. In this case, we define that the center line is the virtual rotation axis representing T0,φ, and rearrange any fixed point on Tr,φ to make one of ideal sinograms. Especially, we translated the fixed point on T0,φ as shown in Fig. 5c.
3D translation error correction
A translation error is the error that occurs when there is a certain movement of the object during the beam time. This movement can be directed to three directions in the 3D space. A compensation for this error can be done through similar way we did with the translation error in 2D space, bringing projected fixed points on the Tr,φ,h function.
where h is a level of a common layer of a projection set. We define a common layer as the plane that envelops the same whole part of the object’s axial level in the real space and perpendicular to the RA.
The translation error in 3D is different from the one in 2D in terms that one should consider layers from where the error occurs. Our numerical results illustrate how the translation error can be compensated by using common layers which include fixed points (Fig. 6). Figure 6(a) shows the sinogram with translation errors (θe = 60°) at the blue colored axial level of the CCD (right panel) and two projection images at θ = 45° and θ = 90°, (left and middle panel, respectively). In order to fix the error, we first set two cleared fixed points in the projection sets at both angles (red and orange colored circles are the first and second fixed points, respectively). However, those fixed points were placed on different axial level of the CCD. For this case, we adjusted the projection images via the high density areas by rearranging the projection set where each high density area is located on the same axial level, called the common layer (Fig. 6(b)). Once this process is completed, the high density area appears its trajectory in the sinogram over θ less than 60°. The shape of the high density area can be considered as a circular shape and the center of the circle can be a projected fixed point. To fix the translation errors, we translate the projected fixed points onto the Tr,φ,h function of the common layer. To make our calculation easier, we place the projected fixed points on the T0,φ,h and then the projected fixed points are located on the projected RA. Adjusting common layers is a significant procedure in the 3-dimensional image reconstruction because it is simply not possible to reconstruct an image well using projections of different layers.
(a) Projection images at θ = 45° (left) and at θ = 90° (middle). Right panel shows the sinogram with errors (θe = 60°) at the blue colored axial level of the CCD. These projection images (at θ = 45° and θ = 90°) contain two cleared fixed points (red and orange circles indicate the first and second fixed points, respectively). (b) Adjusted projection images via the second fixed point. The region between upper and lower red colored lines is a common layer set. A sinogram of each common layer can be converted to the ideal sinogram pattern. The right panel shows the sinogram at the projected common layer containing the second fixed point. The black arrows indicate the trajectory of the second fixed point in the sinogram.
Correction algorithm for the vertical tilt of a RA
In correction process, the priority is to know by which angle the RA is tilted. The whole projected images will be corrected accordingly to the tilted angle using the vertical center line of whole projection as a fiducial line. To calculate the specific angle that the RA is tilted by, we used the FP in this study. Assuming that we virtually collected the every projected fixed point of each projection we get from a specimen and placed them on the CCD, we thought that they would make a virtual line segment just like in Fig. 3b. In fact, the line segment of projected fixed point stands for the trajectory of FP, and the tilted RA is perpendicular to it. If this RA and the vertical line make a certain angle, it is the angle by which we will rotate the projection set to meet the ideal one which doesn’t have the tilt error.
Calculation of tilted angle of RA in parallel tilt errors
Even in a parallel tilt, the tilted angle can be calculated with the same method. The angle of parallel tilt, α, satisfies . Here, a means the major axis and b means the minor axis against the trajectory of CA (Fig. 3c). Whether the orientation of trajectory in the projection is clockwise or counter clockwise depends on whether the azimuthal angle of RA is 0 degree or 180 degrees. This method can be also applied to the mixed error of vertical and parallel tilts. (see Supplementary Fig. S6).
For access to the source code implementing the algorithm, please contact the corresponding author.
Additional Information
How to cite this article: Jun, K. and Yoon, S. Alignment Solution for CT Image Reconstruction using Fixed Point and Virtual Rotation Axis. Sci. Rep. 7, 41218; doi: 10.1038/srep41218 (2017).
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
Bautz, W. & Kalender, W. Radiologe 45, 350–355 (2005).
Kao, T. L. et al. Nanoscale elemental sensitivity study of Nd2Fe14B using absorption correlation tomography. Microsc. Res. Tech. 76, 1112–1117 (2013).
Wang, Y. et al. A high-throughput x-ray microtomography system at the Advanced Photon Source. Rev. Sci. Instrum. 72, 2062–2068 (2001).
Pan, Y., De Carlo, F. & Xiao, X. Automatic detection of rotational centers using GPU from projection data for micro-tomography in synchrotron radiation. Proc. SPIE, 8313, 831328 (2012).
Gürsoy, D., De Carlo, F., Xiao, X. & Jacobsen, C. TomoPy: a framework for the analysis of synchrotron tomographic data. J. Synchrotron Rad. 21, 1188–1193 (2014).
Donath, T., Beckmann, F. & Schreyer, A. Automated determination of the center of rotation in tomography data. J. Opt. Soc. Am. A 23, 1048–1057 (2006).
Parkinson, Y. D., Knoechel, C., Yang, C., Larabell, C. A. & Le Gros, M. A. Automatic alignment and reconstruction of images for soft X-ray tomography. J. Struct. Biol. 177, 259–266 (2012).
Yang, Y. et al. Registration of the rotation axis in X-ray tomography. J. Synchrotron Rad. 22, 452–457 (2015)
Vo, N. T., Drakopoulos, M., Atwood, R. C. & Reinhard, C. Reliable method for calculating the center of rotation in parallel-beam tomography. Opt. Express 22, 19078 (2014).
Hubbell, J. H. Photon mass attenuation and energy-absorption coefficients. Int. J. Appl. Radiat. Isot. 33, 1269–1290 (1982).
Attwood, D. T. Soft X-rays and extreme ultraviolet radiation: principles and applications (Cambridge Univ. Press, New York, 1999).
Le Gros, M. A., McDermott, G. & Larabell, C. A. X-ray tomography of whole cells. Curr. Opin. Struct. Biol. 15, 593–600 (2005).
Weiss, D. et al. Computed tomography of cryogenic biological specimens based on X-ray microscopic images. Ultramicroscopy 84, 185–197 (2000).
Carrascosa, J. L. et al. Cryo-X-ray tomography of vaccinia virus membranes and inner compartments. J. Struct. Biol. 168, 234–239 (2009).
McDermott, G., Le Gros, M. A., Knoechel, C. G., Uchida, M. & Larabell, C. A., Soft X-ray tomography and cryogenic light microscopy: the cool combination in cellular imaging. Trends Cell Biol. 19, 587–595 (2009).
Meyer-Ilse, W. et al. High resolution protein localization using soft X-ray microscopy. J Microsc-Oxford 201, 395–403 (2001).
Schneider, G. et al. Three-dimensional cellular ultrastructure resolved by X-ray microscopy. Nat. Methods 7, 985–U116 (2010).
Uchida, M. et al. Soft X-ray tomography of phenotypic switching and the cellular response to antifungal peptoids in Candida albicans. Proc. Natl. Acad. Sci. USA 106, 19375–19380 (2009).
Uchida, M. et al. Quantitative analysis of yeast internal architecture using soft X-ray tomography. Yeast 28, 227–236 (2011).
Kremer, J. R., Mastronarde, D. N. & McIntosh, J. R. Computer visualization of three-dimensional image data using IMOD. J. Struct. Biol. 116, 71–76 (1996).
Sorzano, C. O. S. et al. XMIPP: a new generation of an open-source image processing package for electron microscopy. J. Struct. Biol. 148, 194–204 (2004).
Amat, F. et al. Markov random field based automatic image alignment for electrontomography. J. Struct. Biol. 161, 260–275 (2008).
Chen, H., Hughes, D. D., Chan, T. A., Sedat, J. W. & Agard, D. A., IVE (Image Visualization Environment): a software platform for all three-dimensional microscopy applications. J. Struct. Biol. 116, 56–60 (1996).
Liu, Y., Penczek, P. A., Mcewen, B. F. & Frank, J. A marker-free alignment method for electron tomography. Ultramicroscopy 58, 393–402 (1995).
Zheng, S. Q. et al. UCSF tomography: an integrated software suite for real-time electron microscopic tomographic data collection, alignment, and reconstruction. J. Struct. Biol. 157, 138–147 (2007).
Acknowledgements
The authors acknowledge the expert help of Kyu Kwon about this manuscript, and Dr. Keith Jones, a senior physicist at Brookhaven National Laboratory, who provided the soil projection sets. This work was sponsored by the Seoul Hayannara Research Center (SHRC).
Author information
Authors and Affiliations
Seoul Hayannara Research Center, Gyeongsangnam-do, Changwon-si, 51495, Korea
Kyungtaek Jun
Department of Dentistry, Seoul National University, Seoul, 03080, Korea
Seokhwan Yoon
Authors
- Kyungtaek Jun
You can also search for this author inPubMed Google Scholar
- Seokhwan Yoon
You can also search for this author inPubMed Google Scholar
Contributions
K.J. conceived and designed the theoretical experiments, and performed the experiment. The theoretical solutions were developed by all authors, and all authors wrote the paper. S.Y. contributed the material.
Corresponding author
Correspondence to Kyungtaek Jun.
Ethics declarations
Competing interests
The authors declare no competing financial interests.
Supplementary information
Rights and permissions
This work is licensed under a Creative Commons Attribution 4.0 International License. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in the credit line; if the material is not included under the Creative Commons license, users will need to obtain permission from the license holder to reproduce the material. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/
About this article
Cite this article
Jun, K., Yoon, S. Alignment Solution for CT Image Reconstruction using Fixed Point and Virtual Rotation Axis. Sci Rep 7, 41218 (2017). https://doi.org/10.1038/srep41218
Received: 01 June 2016
Accepted: 16 December 2016
Published: 25 January 2017
DOI: https://doi.org/10.1038/srep41218