Abstract
Core share and HTML view are not available for this content. However, as you have access to this content, a full PDF is available via the ‘Save PDF’ action button.
We present a brief discussion of sample preparation procedures at the Keck Carbon Cycle Accelerator Mass Spectrometer (KCCAMS), University of California, Irvine, and a systematic investigation of the use of Mg(ClO4)2 as an absorptive water trap, replacing the standard dry ice/ethanol cold finger in graphite sample preparation. We compare high-precision AMS measurement results from oxalic acid I and USGS coal samples using Mg(ClO4)2 under different conditions. The results obtained were also compared with those achieved using the conventional water removal technique. Final results demonstrate that the use of Mg(ClO4)2 as an alternative water trap seems very convenient and reliable, provided the Mg(ClO4)2 is replaced frequently.
References
Loyd, DH, Vogel, JS, Trumbore, S. 1991. Lithium contamination in AMS measurement of 14C. Radiocarbon 33(3):297–301.Google Scholar
Pearson, A, McNichol, AP, Schneider, RJ, von Reden, KF, Zheng, Y. 1998. Microscale AMS 14C measurement at NOSAMS. Radiocarbon 40(1):61–75.Google Scholar
Southon, JR, Santos, GM, Druffel-Rodriguez, K, Druffel, E, Trumbore, SE, Xu, X, Griffin, S, Ali, S, Mazon, M. The Keck Carbon Cycle AMS Laboratory, University of California, Irvine: initial operation and a background surprise. Radiocarbon , .Google Scholar