Feasibility of ambulatory, continuous 24-hour finger arterial pressure recording - PubMed
Feasibility of ambulatory, continuous 24-hour finger arterial pressure recording
B P Imholz et al. Hypertension. 1993 Jan.
Abstract
We tested Portapres, an innovative portable, battery-operated device for the continuous, noninvasive, 24-hour ambulatory measurement of blood pressure in the finger. Portapres is based on Finapres, a stationary device for the measurement of finger arterial pressure. Systems were added to record signals on tape, to alternate measurements between fingers automatically each 30 minutes, and to correct for the hydrostatic height of the hand. We compared the pressure as measured by Portapres with contralateral intrabrachial pressure measured with an Oxford device. Results were obtained in eight volunteers and 16 hypertensive patients. Time lost due to artifact was about 10% for each device. In two patients a full 24-hour Oxford profile was not obtained. In the remaining 22 subjects finger systolic, diastolic, and mean pressures differed +1 (SD 9), -8 (6), and -10 (6) mm Hg, respectively, from intrabrachial pressure. These diastolic and mean pressure underestimations are similar to what was found earlier for Finapres, are typical for the technique, and are systematic. Avoiding brisk hand movements resulted in fewer waveform artifacts. The hand had to be kept covered to continue recording at low outside temperatures. Sleep was not disturbed by Portapres, and arterial pressure showed a marked fall during siesta and nighttime. There were no major limitations in behavior, and no discomfort that originated from continuous monitoring was reported. Measurements continued normally during physical exercise. Portapres provides for the first time continuous 24-hour, noninvasive ambulatory blood pressure waveform monitoring and offers real and obvious advantages over current noninvasive and invasive devices.
Similar articles
-
Schmidt TF, Wittenhaus J, Steinmetz TF, Piccolo P, Lüpsen H. Schmidt TF, et al. J Cardiovasc Pharmacol. 1992;19 Suppl 6:S117-45. doi: 10.1097/00005344-199206001-00019. J Cardiovasc Pharmacol. 1992. PMID: 1382159 Clinical Trial.
-
Estimation of blood pressure variability from 24-hour ambulatory finger blood pressure.
Omboni S, Parati G, Castiglioni P, Di Rienzo M, Imholz BP, Langewouters GJ, Wesseling KH, Mancia G. Omboni S, et al. Hypertension. 1998 Jul;32(1):52-8. doi: 10.1161/01.hyp.32.1.52. Hypertension. 1998. PMID: 9674637
-
Validation of brachial artery pressure reconstruction from finger arterial pressure.
Guelen I, Westerhof BE, van der Sar GL, van Montfrans GA, Kiemeneij F, Wesseling KH, Bos WJ. Guelen I, et al. J Hypertens. 2008 Jul;26(7):1321-7. doi: 10.1097/HJH.0b013e3282fe1d28. J Hypertens. 2008. PMID: 18551006
-
Fifteen years experience with finger arterial pressure monitoring: assessment of the technology.
Imholz BP, Wieling W, van Montfrans GA, Wesseling KH. Imholz BP, et al. Cardiovasc Res. 1998 Jun;38(3):605-16. doi: 10.1016/s0008-6363(98)00067-4. Cardiovasc Res. 1998. PMID: 9747429 Review.
-
Kim SH, Lilot M, Sidhu KS, Rinehart J, Yu Z, Canales C, Cannesson M. Kim SH, et al. Anesthesiology. 2014 May;120(5):1080-97. doi: 10.1097/ALN.0000000000000226. Anesthesiology. 2014. PMID: 24637618 Review.
Cited by
-
Faria AP, Ritter AMV, Santa-Catharina A, Souza DP, Naseri EP, Bertolo MB, Pioli MR, Carvalho CC, Modolo R, Moreno H. Faria AP, et al. Arq Bras Cardiol. 2021 Mar;116(3):443-451. doi: 10.36660/abc.202190703. Arq Bras Cardiol. 2021. PMID: 33909773 Free PMC article. Clinical Trial. English, Portuguese.
-
Houtman S, Oeseburg B, Hughson RL, Hopman MT. Houtman S, et al. Clin Auton Res. 2000 Aug;10(4):207-12. doi: 10.1007/BF02291358. Clin Auton Res. 2000. PMID: 11029019 Clinical Trial.
-
Konstantinidis D, Iliakis P, Tatakis F, Thomopoulos K, Dimitriadis K, Tousoulis D, Tsioufis K. Konstantinidis D, et al. J Hum Hypertens. 2022 Nov;36(11):945-951. doi: 10.1038/s41371-022-00675-z. Epub 2022 Mar 23. J Hum Hypertens. 2022. PMID: 35322181 Free PMC article. Review.
-
Parati G, Castiglioni P, Faini A, Di Rienzo M, Mancia G, Barbieri R, Saul JP. Parati G, et al. Front Physiol. 2019 May 7;10:477. doi: 10.3389/fphys.2019.00477. eCollection 2019. Front Physiol. 2019. PMID: 31133867 Free PMC article.
-
Sajgalik P, Kremen V, Carlson AR, Fabian V, Kim CH, Wheatley C, Gerla V, Schirger JA, Olson TP, Johnson BD. Sajgalik P, et al. J Appl Physiol (1985). 2016 Dec 1;121(6):1319-1325. doi: 10.1152/japplphysiol.00981.2015. Epub 2016 Oct 20. J Appl Physiol (1985). 2016. PMID: 27765846 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources